We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Purpose: To determine the feasibility of Ultra wide-field intravenous fundus fluorescein angiography (UWF-IVFFA) in infants. Methods: A retrospective review of infants who underwent UWF-IV-FFA with the Optos P200MA was performed. The ability to capture different phases of the angiogram, the total image acquisition time and the presence of image artefacts were analysed. Results: Twelve infants with a variety of proliferative retinopathies underwent UWF-IV-FFA over a 7-month period. The mean age was 3.4 months (range 2-6 months) and the mean image acquisition time was 4 min (range 3-5). Pseudocolour fundus images and the venous, recirculation and late phases of UWF-IV-FFA were captured successfully in all infants (100%). Choroidal and arterial phases were captured in one (8.3%) and 10 (83.3%) infants, respectively. Image artefacts due to eyelashes and corneal desiccation occurred in nine (75%) and six (50%) infants, respectively. No adverse medical problems were noted. Conclusions: We have shown that UWF-IV-FFA can be successfully and safely performed in infants using the Optos P200MA system. It is a feasible new alternative to RetCam for obtaining an IV-FFA in the outpatients setting.

Original publication

DOI

10.1136/bjophthalmol-2013-304450

Type

Journal article

Journal

British Journal of Ophthalmology

Publication Date

01/03/2014

Volume

98

Pages

302 - 304